OEM/ODM
Profession
Certificate
Open-Close Cycles>1kTimes
Close cycles using the closures on the test specimens.
Leak Rate<4mmHg/min
Acceptance Criteria: Air Leakage shall not cause a pressure drop that exceeds 4mmHg/min.
Life Test>30kTimes
Blood pressure cuffs alone will be subject to 30000 inflation cycles from 150 mmHg to 300mmHg.